Singh, RM and Saini, PK and Mathur, SC and Singh, GN and Lal, B (2010) Development and validation of a RP-HPLC method for estimation of montelukast sodium in bulk and in tablet dosage form. Indian Journal of Pharmaceutical Sciences, 72 (2). p. 235. ISSN 0250-474X
Full text not available from this repository.
Official URL: https://doi.org/10.4103/0250-474X.65023
Abstract
HPLC method developed for estimation of montelukast sodium in bulk and tablets using C18 column and acetonitrile:sodium acetate (90:10 v/v, pH 6.3). Flow rate 1.5 ml/min, detection at 285 nm, retention time 3.4 min. LOD 1.31 μg/ml, LOQ 3.97 μg/ml. Validated for linearity, precision, accuracy, specificity. Suitable for routine quality control. © 2010 Elsevier B.V., All rights reserved.